Care and Maintenance
10-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
10-7-5-2 Data Sheet for enclosure Source Leakage Current
The test passes when all readings measure less than the value
shown in Table 10-12 on page 10-21. Record all data on the PM
Inspection Certificate.
Table 10-16: Typical Data Sheet for enclosure Source Leakage Current
Unit
Power
Tester
Polarity
Switch
Tester
Neutral
or
Ground
Switch
Test 1
Speaker
Cover
Test 2 Real
Panel Metal
Parts
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED