PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
14-15
14- 8 SG120 Wall Stand (GPCP No.: 2402562) Specifications
See Table 14-26, 14-27.
TABLE 14-26
SG120 WALL STAND SPECIFICATIONS
Parameter Specification
Weight
220kg (485.1 lbs)
Height
2235 mm (87.99 in)
Min. Width
652 mm (25.67 in)
Max. Width
915 mm (36.02 in)
Min. Depth
687 mm (27.05 in)
Max. Depth
927 mm (36.5 in)
Front panel height
64.2 cm (25.2 in)
Front panel width
65 cm (25.6 in)
Distance Front Panel to Film Plane
5.8 cm (2.28 in)
Min. Height (Bucky vertical)
33 cm (13.1 in)
Max. Height (Bucky vertical)
190.5 cm (75 in)
Min. Height (Bucky horizontal)
63.5 cm (25 in)
Max. Height (Bucky horizontal)
213 cm (83.9 in)
Bucky rotation range
0° to 180°
Bucky Angulation range
–20° to 90°.
Grid Holding Force
>196N
Wall stand panel Filtration
Maximum 0.65 Al equivalency at 100 kVp
Bucky Loading
Left hand or Right hand
Cassette Load
Left Hand (Right hand load field configurable)
Power Supply
Single phase voltage: 115, 230V+/-10%+ground
Line Frequency
50/60 Hz
Max. Current (Moment.)
0.4
Max. Current (Contin.)
0.2
Panel to front column distance
550 mm (21.65 in)
Panel to front cover distance
483 mm (19.01 in)
TABLE 14-27
SG120 WALL STAND BUCKY SPECIFICATIONS
Parameter Specification
Grid
36L/cm; 130/180 cm focus
Cassette Size Sensing
13 cm×18 cm (5 in.×7 in.) minimum to 35 cm×43 cm
(14 in.×17 in.) maximum
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).