Page #247 background image
Personal Information Collected by the Product
Vscan ExtendUser Manual 8-21
5721203-100 Rev. D
Potential Hazardous Situations Resulting From Failures of the IT
Network
Hazardous situations
The following situations have been identified as potentially
hazardous as a result of the IT network failing to provide the
required characteristics specified above.
1. Delayed or impaired access to images or other exam
information or patient data.
2. Permanent loss of images or other exam information or
patient data.
3. Corruption of images or other exam information or patient
data.
Warning
In addition to the hazardous situations identified above,
connection of the Vscan Extend to a network that includes other
equipment could result in other unidentified risks to patients,
operators or third parties. The responsible organization should
identify, analyze, evaluate and control these risks on an ongoing
basis. Re-evaluate the risks if any of the following occurs:
1. Changes in network configuration
2. Connection of additional items to the network
3. Disconnecting items from the network
4. Update of equipment connected to the network
5. Upgrade of equipment connected to the network