EasyManuals Logo
Home>GE>Medical Equipment>Vscan Extend

GE Vscan Extend User Manual

GE Vscan Extend
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
Vscan ExtendUser Manual i-5
5721203-100 Rev. D
Classifications
The following classifications are in accordance with the IEC/
EN 60601-1.
Type and degree of protection against electric shock:
Vscan Extend has an internal battery which allows the
operation during AC power absence.
The AC adapter is Class II.
Vscan Extend has type BF Applied Part.
Vscan Extend main unit is rated IP33:
3: Protected against solid foreign objects of 2,5 mm Ø and
greater.
3: Protected against spraying water.
Vscan Extend probe (immersible portion) is IPX7.
Class II Equipment
EQUIPMENT in which protection against electric shock does not
rely on BASIC INSULATION only, but in which additional safety
precautions such as DOUBLE INSULATION or REINFORCED
INSULATION are provided, there being no provision for
protective earthing or reliance upon installation conditions.
Type BF Applied part
TYPE BF APPLIED PART providing a specified degree of
protection against electric shock, with particular regard to
allowable LEAKAGE CURRENT.
Original Documentation
The original document was written in English.
Table i-2: Leakage Current
Normal mode Single fault condition
Patient leakage current <100 microA <500 microA