4
Maintenance and inspection procedures
Mechanical inspections
20 / 44
Lucea 50_100
01740 EN 01
Adjusting the top stop
5 mm
Fig.15: Top stop adjustment
Tighten to lower the stop.
Loosen to raise the stop.
4.4 Mechanical inspections
4.4.1 Visual inspections
Fig.16: Integrity of configuration
Check the entire surgical light (spring arm,
complete suspension with ceiling cover)
for signs of damage
Fig.17: Checking the verticality
Check that the suspension tube is vertical
(ceiling-mounted version).