4
Maintenance and inspection procedures
Hydraulic inspections
30 / 44
Lucea 50_100
01740 EN 01
Check Mains LED 1 Battery LED 2 LEDs3 to8 3 Meaning
After 1 hour
(LCA50) or 4
hours (LCA100)
Off Yellow One of LEDs
3to8 is lit (bat-
tery charge level)
Operation on bat-
teries
Connect the
power outlet
Green Off Scrolling LEDs Batteries char-
ging
Tab.7: Battery lifetime test
4.5.2 Visual inspections
Fig.39: Remote control
Remote control (option)
1. Check that the remote control operates
correctly.
2. Check the state of the batteries.
3. Check the lighthead selection function.
The video camera compatible with the
LUCEA100 lightheads is no longer available
since January2019.
Fig.40: Power lead for mobile version
Power lead (mobile version only)
1. Check that the power lead is not dam-
aged.
2. Check that the IEC mains connector on
the power supply enclosure cover is cor-
rectly connected
4.6 Hydraulic inspections
Not applicable to this product