Servo-i Ventilator System Disassembling and assembling
Revision 02 Service Manual 4 - 17
4
Carefully release and remove the retaining ring
pliers.
Check that the membrane is not deformed by the
retaining ring. If necessary, remove the membrane
and redo the complete mounting procedure.
Mount the expiratory cassette onto the Patient
Unit.
Reset the operating capacity meter after
replacement of the membrane. To access the reset
button, select Menu / Biomed / Service.
Note: If the metal washer has been separated from
the membrane, it is important that the washer is
correctly mounted. The washers raised hub (1) must
be facing outwards as shown in the adjacent
illustration.
Squeeze the retaining ring pliers and carefully
press the membrane in place into the cassette as
shown in the illustration.
SVX9059SVX9060SVX9058