Disassembling and assembling Servo-i Ventilator System
4 - 18 Service Manual Revision 02
4
Control cable
The control cable connectors must be connected
carefully to avoid damages on the connector pins.
• Carefully find the correct position; the connector
pins and the guides in the connectors must
correspond. Do not turn the connector while
inserting!
• When correct position is found, insert the
connector.
• Secure the connector with its locking ring.
The illustration shows the Patient Unit connector, but
the procedure for the User Interface connector is the
same.
SVX9061
Fixed battery module
If the upper slot in the Module unit is equipped with a
latch, a tool and two-hand operation is required to
remove the Battery module. To release the fixed
battery module:
1. Push a suitable tool, e.g. a small hexagonal
wrench or a screwdriver, through the hole as
shown in the illustration.
2. At the same time, push the release button to the
right.
3. Remove the tool.
4. Continue to push the release button until the
Battery module snaps out.
5. Pull out the Battery module.
SVX9085
SVX9086