11
The Control Solutions
Read carefully the instruction for use of the control solution that
you want to use.
It is possible to check the working of the multiCare-in instrument
using the multiCare-in control solutions.
The control solutions can be obtained as a separate accessory
through specialized retailers.
The control solution is applied on the strip just like blood.
When testing is done with the control solution, the obtained result
must be within the interval as shown on the label of the strip’s vial.
In this way, it is possible to learn how to use the multiCare-in properly
and at the same time, check if it is working.
Write the date of opening on the vial when the solution is opened for
the rst time.
The control solution must be used within 90 days from rst opening,
if the expiry date is valid.
When to carry out a working check up using the control solutions:
• When suspecting a possible defect of the multiCare-in instrument.
• When suspecting the possibility of a malfunction or wrong
measurement.
The result obtained with the control solutions has to be deleted by
the memory of the meter in order to not interfere with the average
calculation of the glycaemia.