Figure 1 – Setting the bath for heat transfer beads
When operating the bath configured for heat transfer beads the temperature range is
limited to 80°C maximum. The Dry Protection Alarm is no longer required so is not
available on the bath menu.
Important note. Baths using heat transfer beads behave differently from water baths
The performance specification of Grant baths detailed in Section 9 is for water only.
Performance with heat transfer beads will be significantly different. Users should refer to
the bead manufacturer’s recommendations for use and performance.
Any existing protocols for sample preparation will need to be revalidated to prevent
possible overheating.
It is recommended that a drain stopper be used to prevent beads entering the drain hole.
These are available to purchase in a pack of five, part number UWB-DS.
Select the menu option
Navigate to the Heat transfer
selection menu option
Press set to select the menu.
The water setting is
displayed
Press the up button to select
the option
Press the set key to save the option.
The bath with automatically return to
displaying the actual temperature.