9 Technical specifications
Operating conditions
Ambient Temperature 5 to 40°C
Maximum relative humidity 80% R.H. in room temperatures up to 31°C
decreasing linearly to 50 % R.H. at 40°C
Altitude above sea level
Up to 2,000 m (6,500 ft)
Operating Environment
Indoor use only
Electrical details
Mains supply: 220-230V @ 50/60 Hz or 110-120V @ 50/60 Hz
Pollution degree: 2
Installation Category: II
Note: Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply
voltage
Models Capacity (L)
Current rating (A)
120V
230V
SAP2
2
2.0
1.0
SAP2S
2
3.0
1.5
SAP5 5 3.0 1.5
SAP12
12
6.5
3.5
SAP18 18 8.5 6.0
SAP26
26
8.5
6.0
SAPD 5 & 12 9.5 5.0
SAP34
34
10.5
8.0
JBN5
5
3.0
1.5
JBN12 12 6.5 3.5
JBN18
18
8.5
6.0
JBN26 26 8.5 6.0
JBA5
5
3.0
1.5
JBA12 12 6.5 3.5
JBA18
18
8.5
6.0
SUB Aqua Pro bath performance
Temperature range
5°C above ambient to 99°C
Display (also used for setting)
10.0 to 99.0°C in 0.1°C steps
Temperature stability ± 0.2°C
JB Academy & JB Nova bath performance
Temperature range
5°C above ambient to 95°C
Setting scale
10 to 95°C in 0.5°C steps
Temperature stability ± 0.5°C
All performance data specified tested in accordance with DIN12876.

Table of Contents

Other manuals for Grant SAP12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Grant SAP12 and is the answer not in the manual?

Grant SAP12 Specifications

General IconGeneral
BrandGrant
ModelSAP12
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals