Error Messages & Messages
The following error messages or messages may appear on the ASR-
2000 screen display during varied operations.
Positive output current peak value
is excessive. Press "Shift + Cancel"
to clear this alarm.
Negative output current peak
value is excessive. Press "Shift +
Cancel" to clear this alarm.
Internal power stage over heat.
Press "Shift + Cancel" to clear this
alarm. Check if operating
temperature is suitable and if filter
is clogged.
Output current RMS value is
excessive. Press "Irms" to check
allowance set range
The power input voltage is
insufficient or turning off main
power switch. Check input power
before rebooting the unit.
Fan failure. Contact service center.
Over internal power stage
maximum power (110% of rating
power), press "Shift + Cancel" to
clear this alarm.
Call attention to output terminal
short status
Over internal maximum voltage
(110% of rating voltage). Press
"Shift + Cancel" to clear this
alarm.
The calibration data is abnormal
or out of allowance range. Contact
service center.