3-71.
Remote/Local
Capability:
RL1.
The
4262A
goes
to
Remote
Status
only
when
it
accepts
Listen
address
with
REN
(Remote
Enable)
line
in
the
Bus
lines
set
to
“1”’.
Remote
status
is
not
obtained
if
REN
line
is
set
to
““1”’
after
Listen
address
is
received.
Remote
status
is
returned
to
Local
status
when
one
of
following
conditions
is
present:
(1)
REN
line
is
set
to
“0”.
(2)
LOCAL
switch
on
front
panel
is
depressed.
(3)
GTL
(Go
To
Local)
command
is
received.
Local
Lock
Out:
LLO
Local
Lock
Out
inhibits
the
function
of
LOCAL
switch,
This
LLO
command
is
a
universal
command
and
is
valid
when
REN
line
is
set
to
“1”.
LLO
command
is
disabled
when
REN
line
is
set
to
“0”
3-72.
Device
Clear
Capability:
DC1.
The
4262A
is
set
to
initial
conditions
(the
same
conditions
that
are
automatically
set
after
turn-on
of
power
switch),
when
it
accepts
DCL
(Device
CLear)
command—universal
command—or
SDC
(Selected
Device
Clear)—addressed
command.
3-73.
Device
Trigger
Capability:
DT1.
The
4262A
is
triggered
regardless
of
TRIGGER
switch
setting
when
it
accepts
GET
command-ad-
dress
command.
3-74.
ADDRESS
Switch:
ADDRESS
switch
on
the
rear
panel
sets
Listen/Talk
address.
Five
section
or
five
bit
switch
provides
30
settings
from
00000
to
11110.
A5
A4
A3 A2
Al
0 0 0
0
0
ooee
0
3-75.
Remote
Message
Coding:
Interface
'Bus
Command
signals
for
the
4262A
are
listed
in
Table
3-61.
3-36
Model
4262A

Table of Contents

Other manuals for HP 4262A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 4262A and is the answer not in the manual?

HP 4262A Specifications

General IconGeneral
BrandHP
Model4262A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals