Model
4262A
Section
IV
Paragraphs
4-1
to
4-8
SECTION
1V
PERFORMANCE
TESTS
4-1.
INTRODUCTION.
4-2.
This
section
provides
the
check
procedures
to
verify
the
4262A
specifications
listed
in
Table
1-1.
All tests
can
be
performed
without
ac-
cess
to
the
interior
of
the
instrument.
A
simpler
operational
test
is
presented
in
Section
III
under
Self
Test
(paragraph
3-5).
The
performance
test
procedures
in
this
section
can
also
be
used
to
do
an
incoming
inspection
of
the
instrument
and
to
verify
whether
the
instrument
meets
its
speci-
fied
performance
after
troubleshooting
or
making
adjustments.
If
specifications
are
found
to
be
out
of
limits,
check
that
controls
are
properly
set,
and
then
proceed
to
adjustments
or
troubleshooting.
Note
Allow
a
15-minute
warm-up
and
stabilization
period
before
conduct-
ing
any
performance
test.
4-3.
EQUIPMENT
REQUIRED.
4-4.
Equipment
required
for
the
performance
tests
is
listed
in
Table
1-4
Recommended
Test
Equip-
ment
in
Section
I.
Any
equipment
whose
characteristics
equal
the
critical
specifications
given
in
the
table
may
be
substituted
for
the
recom-
mended
model(s).
Accuracy
checks
in
this
section
use
standard
LCR
components
as
the
samples
to
be
connected
to
the
4262A.
Accessories
16361A
and
16362A
can
be
utilized
for
this
purpose.
These
accessory
models
are
DUT
(device
under
test)
boxes
from
which
the
desired
component
can
be
selected
and
connected
to
the
4262A
through
cables
by
use
of
a
rotary
switch.
If
models
16361A/16362A
are
unavailable,
use
the
discrete
components
rec-
ommended
in
Table
4-1.
Note
All
components
used
as
standards
should
be
calibrated
by
an
instru-
ment
whose
specifications
are
traceable
to
NBS,
PTB,
LNE,
NRC,
JEMIC,
or
equivalent
standards
group;
or
all
components
should
be
calibrated
directly
by
an
authorized
calibration
organization
such
as
NBS.
The
calibration
cycle
should
be
determined
by
the
stability
spec-
ification
for
each
component.
45.
TEST
RECORD.
4-6.
Results
of
the
performance
tests
may
be
tabu-
lated
on
the
Test
Record
at
the
end
of
these
proce-
dures.
The
Test
Record
lists
all
the
tested
specifica-
tions
and
their
acceptable
limits.
Test
results
re-
corded
at
incoming
inspection
can
be
used
for
comparison
in
periodic
maintenance
and
trouble-
shooting
and
after
repairs
or
adjustments.
4-7.
CALIBRATION
CYCLE.
4-8.
This
instrument
requires
periodic
verification
of
performance.
Depending
on
the
use
and
envir-
onmental
conditions,
the
instrument
should
be
checked
with
the
following
performance
tests
at
least
once
every
year.
To
maximize
the
“up
time”
of
the
instrument,
the
recommended
preventive
maintenance
frequency
for
the
4262A
is
twice
a
year.
4-1

Table of Contents

Other manuals for HP 4262A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 4262A and is the answer not in the manual?

HP 4262A Specifications

General IconGeneral
BrandHP
Model4262A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals