8-19
Preset State and Memory Allocation
Memory Allocation
If correction is on at the time of an external store, the calibration set is stored to disk.
(Note that inactive calibrations are not stored to disk.) When an instrument state is loaded
into the analyzer from disk, the stimulus and response parameters are restored first. If
correction is on for the loaded state, the analyzer will load a calibration set from disk that
carries the same title as the one stored for the instrument state.
Conserving Memory
If you are concerned about conserving memory, either internal memory or external disk
space, some of the most memory-intensive operations include:
two-port error correction
interpolated error correction
1601 measurement points
using time domain
saving data arrays and graphics with the instrument state
Using Saved Calibration Sets
When you are saving to internal memory (CMOS, non-volatile memory), calibration sets
are linked to the instrument state and measurement parameter for which the calibration
was done. Therefore a saved calibration can be used for multiple instrument states as long
as the measurement parameter, frequency range, and number of points are the same. A
full 2-port calibration is valid for any S-parameter measurement with the same frequency
range and number of points. When an instrument state is deleted from memory, the
associated calibration set is also deleted if it is unused by any other state.
The following hints will help you avoid potential problems:
If a measurement is saved with calibration and interpolated calibration on, it will be
restored with interpolated calibration on.
A calibration stored from one instrument and recalled by a different one will be invalid.
To ensure maximum accuracy, always recalibrate in these circumstances.
No record is kept in memory of the temperature when a calibration set was stored.
Instrument characteristics change as a function of temperature, and a calibration
stored at one temperature may be inaccurate if recalled and used at a different
temperature. Refer to Chapter 1 , “HP 8719/20/22ES Specifications and
Characteristics, andChapter 2 ,“HP 8719/20/22ETSpecifications andCharacteristics,
for allowable temperature ranges for individual specifications.

Table of Contents

Other manuals for HP 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8719ET and is the answer not in the manual?

HP 8719ET Specifications

General IconGeneral
BrandHP
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals