HP 8719/20/22ES Specifications and Characteristics
Instrument Specifications
Power Meter Calibration Accuracy
Table 1-46 Power Meter Calibration Sweep Speed and Accuracy
Power Desired at
Test Port
Number of
Readings
Sweep Time
Setting (seconds)
a
a. Sweep speed applies to every sweep in continuous correction mode, and to the first
sweep in sample-and-sweep mode. Subsequent sweeps in sample-and-sweep mode
will be much faster.
Characteristic
Accuracy (dB)
b
b. The accuracy values were derived by combining the accuracy of the power meter and
linearity of the analyzer's internal source, as well as the mismatch uncertainty
associated with the power sensor.
+5 dBm
1
2
3
33
64
95
±0.7
±0.2
±0.1
15 dBm
1
2
3
48
92
123
±0.7
±0.2
±0.1
30 dBm
1
2
3
194
360
447
±0.7
±0.2
±0.1