12
EDS 3000 Operating Manual
PN#02080105 • ELC1603-1752 / 09.16
ELECTRONICS
Setting Display Setting Range Pre-Setting
Analog Output
MAMP /
VAMP
MAMP
The analog output provides a 4
to 20mA DC signal.
The analog output provides a 0
to 10VDC signal.
Zero Point Calibration (Calibrate)
YES / NO NO
The actual pressure is saved
as the new zero point. This is
possible in the range +/- 3%
of the unit’s nominal pressure.
This will appear calibration is
carried out in the permissible
range, otherwise
is displayed
This function is useful, for example, if there
is always a residual pressure in the system
which should be displayed as 0 bar.
Warning: Following a zero point adjust-
ment, for example, on a 600 bar unit, a
pressure up to 18 bar will be displayed as
0 bar. Before any work is carried out on the
hydraulic system, ensure that the system is
pressurized.
Version Number
Displays the current software version.
(for reference only)
End
YES / NO NO
This close the basic settings
menu
Basic settings can still be
altered
If menu adjustments have been altered, “ProG” appears briefly when quitting the
basic settings menu, then the selected display value is shown in the primary display.