GPIB 488.1 INTERFACE FUNCTIONS
Interface Function
Subset
Description
Source Handshake
SH1
Complete Source handshake capability
Acceptor Handshake
AH1
Complete Acceptor handshake capability
Talker
T6
Talker functions (unaddress if MLA)
Listener
L4
Listener functions (unaddress if MTA)
Service Request
SR1
Complete Service request capability
Remote Local
RL0
No remote/local capability
Parallel Poll
PP0
No parallel poll capability
Device Clear
DC1
Complete Device clear capability
Device Trigger
DT0
No device trigger capability
Controller
C0
No controller capability
Electrical Interface
E2
Three-state drivers
Controllable Items
Test and Reset control.
Setting of test parameters for tests.
Reading of instrument status and test results.
Data Codes
ASCII
Delimiter
NL (+ EOI)
8.4 USB/GPIB/RS-232 Interface Command List
A GPIB read command must be sent after the command strings, to retrieve any
data from a query command (?). The APT 400XAC series GPIB bus will not send
any data to the controller without being queried. The USB/RS-232 bus will
automatically send any response back to the controller’s input buffer. Each
command string should be terminated the ASCII control code, New Line <NL>,
OAh or the end of line EOL message for GPIB.
The following conventions are used to describe the commands syntax. Braces
({ }) enclose each parameter for a command string. Triangle brackets (< >)
indicate that you must substitute a value for the enclosed parameter. The Pipe ( | )
is used to separate different parameter options for a command. Do not include
any of the above characters when sending the commands. The command and
the value should be separated with a space.
All commands that end with a question mark (?) are query commands and
required an IEEE-488 read command to retrieve the data from the device’s
output buffer.
8.4.1 Basic Commands and Query Commands
The following commands are used to control actual output voltage and current