D-0120588-B – 2019/03
Eclipse – Instructions for Use - EN Page 57
3.13.3 Summary of parameters for P300 and MMN
P1, N1, P2 suprathreshold P300
Awake and quit adults, children
and infants
Awake and quit adults, children
and infants
Attend or ignore conditions
Tone burst, speech vowels or
consonant vowel combinations
Tone burst, speech vowels or
consonant vowel combinations
50-300ms
Be careful of overlapping
response if analysis time is short
Oddball paradigm
Deviant probability 0.05-0.20
Number of deviants at least 200
Tip of nose of averaged reference
(jumped electrodes)
Tip of nose of averaged
reference (jumped electrodes)
Pre stimuli -100ms
Post stimuli 700ms or more
Pre stimuli -50ms or more
Post stimuli 400ms or more
At least 2, resulting in at least
200 deviants.
Adult
Children
Infants
Measures
P1, N1, P2
P1, N200-250
Reliable components
Baseline to peak amplitude, peak
latency
Use latency window established
using grand mean data
Any age, use difference
waveform (response to deviant)
Baseline to peak amplitude, peak
latency
Consider mean MMN amplitude
in response window
Use latency window established
using grand mean data
Replicable components
Response 2-3x larger than
amplitude in prestimulus interval
Replicable components
Response 2-3 times larger than
amplitude in prestimulus interval
Further information about the P300 & MMN recordings can be obtained from Eclipse Additional Information
Manual.