SERVICE MANUAL
Invacare
®
- Leo
16
5 Operational Faults
5.1 Operational faults
Please proceed as follows if you have any problems:
• First assess the possible cause of the problem using the following table.
• Check the drive console status display. Evaluate the blink error code.
• Carry out the necessary checks and repairs as recommended in the following table.
5.1.1 Drive fault diagnosis
OTHER
SYMPTOMS
POSSIBLE
CAUSE
SOLUTION Documentation
not start
Status display
on operating
console does
not illuminate
batteries
defective
batteries
See operating manual
completely
discharge battery
batteries
Power supply to
possibly
interrupted
fuse
See operating
instructions for main fuse
position
between
modules for
loose
connections or
See "Replacing the
electronics" on page 40
Operating
console defective
operating
console
See "Replacing
operating console" on
page 36
Operating unit
status display
blinking
Various causes
code
See "Error Codes and
Diagnostic Codes" on
page 17