Remote sense function
Remote sense can adjusted at the output voltage of the device under test, this feature
allows to compensate the voltage drop on the wire between the front panel terminals of the
power supply and the device under test.
Use local sense:
Local sense doesn’t compensate the voltage drop on the connection wire, the operation is:
1. Use the short clips on the back panel of the instrument, or install wire between Vo+ an
d
Vs
+ 、Vo- and
Vs-
2.
Connect the the positive and negative terminals of the front panel to the device
under
test
Use remote sense:
1. Disconnect the wires/short clips between Vo+ and Vs+、 Vo- and Vs-
2.
Connect
wires from Vs+ 、Vs- to the device usder te
st
3. Connect
wires from Vo+ 、Vo- to the device under
test
Note:
In order to ensure the stability of the system, using armored twisted pair cable
between the remote sense terminal of IT6700 and load.
Please note that the positive and negative polarity when wiring, otherwise it
will damage the instrument!
Vo+ Vs+ Vs- Vo-