Calibration Commands
Copyright © Itech Electronic Co., Ltd. 53
CALibrate:VOLTage [:DATa] {<numeric value>}
This command returns the actual output voltage of the current calibration point
of the power supply.
Command syntax
CALibrate:VOLTage [:DATa] <NRf>
Parameters
<NRf>
Example
CAL:VOLT 30.0002V
CALibrate:CURRent:LEVel
This command is used to specify the current calibration point. P1, P2 calibration
points must be sequentially calibrated.
Command syntax
CALibrate:CURRent:LEVel <point>
Parameters
P1|P2
CALibrate:CURRent [:DATa] {<numeric value>}
This command returns the actual output current value of the current calibration
point of the power supply.
Command syntax
CALibrate:CURRent [:DATa] <NRf>
Parameters
<NRf>
Example
CAL:CURR 3.0002A
CALibrate:POWer:LEVel
This command is used to specify the power calibration point. P1, P2 calibration
points must be sequentially calibrated.
Command syntax
CALibrate:POWer:LEVel <point>