Functions and Characteristics
Copyright © Itech Electronic Co., Ltd. 39
3.10.6 Measurement of Voltage Rise Time
The IT8500+ series electronic load is provided with special voltage rise/drop
time measurement function. This function gives a simple analog of voltage
rise/drop speed of oscilloscope test power.
Operation methods:
Set initial Voltage and Final Voltage
1. Press [Shift] + [9] keys to enter configuration menu. Press Right key.
Select Measure and press [Enter] key.
2. Press to select TIMEV1. Press [Enter] key. Press numeric keys
to set initial voltage value and press [Enter] key.
3. Press to select TIMEV2. Press [Enter] key. Press numeric keys
to set final voltage value and press [Enter] key.
4. Press [ESC] to exit setting.
Start timer function
5. Press [Shift] +[8] keys to enter system menu. Press Right key till
DISPLAY” flicks and press [Enter] key.
6. Press key to select “On”. Start timer function and press [Enter]
key.
7. Press [ESC] to exit setting.
8. VFD second line will display time 0.0000S.
0.0001V 0.0002A
0.0000S I= 0.000A
OFF CC
Measurement of Rise Time
9. Connect DC power to be tested to the input terminal of the electronic load.
The power is set with a value that is higher than the set final voltage value.
Keep power output in OFF status.
10. Set a constant current value on the load and open the load input.
11. Open the power output.
12. The electronic load timer starts timing. After ending, time will keep stable,
which is rise time of voltage.
13. Turn off the power output. The electronic load VFD will display voltage drop
time.
3.11 Saving and Recalling Settings
We can save some often-used parameters in the non-volatile memory,
including working mode, voltage/current value and so on. IT8500plus series
provide 100 non-volatile registers.
They are divided into 10 Memory groups: Group0-9.You can set it in the system
menu.
Group0 means you can save and recall parameters in 0-10 registers.
Group1 means you can save and recall parameters in 11-20 registers.
Group2-Group9 can be concluded in the same manner.

Table of Contents

Other manuals for ITech IT8500plus Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8500plus Series and is the answer not in the manual?

ITech IT8500plus Series Specifications

General IconGeneral
BrandITech
ModelIT8500plus Series
CategoryTest Equipment
LanguageEnglish

Related product manuals