Measurement Commands
Copyright © Itech Electronic Co., Ltd. 9
Chapter2 Measurement Commands
MEASure
FETCh
This command is used to read the measurement value of electronic load.
The sequence of measurement data is listed in the following table:
Measurement Value Description
CURR_DC Direct current value
CURR_RMS Current effective value
CURR_MAX Current maximum value
CURR_MAXP The positive peak current
CURR_MINP The negative peak current
VOLT_DC Direct voltage value
VOLT_RMS Voltage effective value
VOLT_MAX Maximum voltage
POW_ACT Active power
POW_APP Apparent power
POW_REAC Reactive power
POW_MAX Maximum power
RES Resistance
FREQ Frequency
CURR_CFAC Crest Factor
POW_PFAC Peak Factor
VOLT_THD Total harmonic distortion of voltage
ETIME Elapsed time under timing mode
TEMP Temperature
Returned value:
<NR3>
MEASure:CURRent?
FETCh:CURRent?
This command is used to read the current average value in load input terminal.
Command syntax:
MEASure[:SCALar]:CURRent[:DC]?
FETCh[:SCALar]:CURRent[:DC]?

Table of Contents

Other manuals for ITech IT8625

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8625 and is the answer not in the manual?

ITech IT8625 Specifications

General IconGeneral
BrandITech
ModelIT8625
CategoryTest Equipment
LanguageEnglish

Related product manuals