Common Commands
Copyright © Itech Electronic Co., Ltd. 59
Chapter7 Common Commands
*CLS
This command is used to clear following registers:
Standard Event Register
Questionable Event Register
Status Byte Register
Command syntax
*CLS
Parameter
None
*ESE
This command is used to edit the value of standard event enable register. It
defines the specified bits from standard event register that will cause the value
of ESB bit in status byte register to be 1.
Command syntax
*ESE <NR1>
Parameter
0~255
Power-On Value
Refer to *PSC command
Example
*ESE 128
Query Syntax
*ESE?
Returned value
<NR1>
Relevant Command
*ESR? *PSC *STB?