Calibration Commands
Copyright © Itech Electronic Co., Ltd. 55
Parameter:
External reading
Unit:
V (volts)
Example:CAL:VOLT 3.2223
Relevant Command:
CAL:STAT CAL:SAV
CALibrate:CURRent:POINt
This command is used to specify the current calibration points.
The calibration points P1, P2, P3 and P4 must be calibrated in order.
Command syntax:
CALibrate:CURRent:POINt <point>
Parameter:
P1 | P2 | P3 | P4
Example:CAL:CURR:POIN P2
Relevant command:
CAL:STAT CAL:SAV
CALibrate:CURRent
This command is used to enter a calibration current value that you obtain by
reading an external meter.
Command syntax:
CALibration:CURRent[:LEVel] <NRf>
Parameter:
External reading
CALibrate:VOLTage:METer:POINt
This command is used to calibrate the measurement voltage point.
Command syntax:
CALibrate:VOLTage:METer:POINt <point>