JWI-700C 16
9. Troubleshooting and Error message
Error message Problems Solutions
Initial zero point exceeds
/-
30% (take 30% as
reference basis).
1. To check whether there a
re other alien
articles on the scale pan, remove those
articles.
2. LOAD CELL failure, which requires to be
changed or to contact our Service.
Higher or lower than A/D
resolution range.
1. Check whether it is A/D failure, if yes,
please replace AD.
2. LOAD CELL failure, replacement is
required or contact our Service.
EEPROM Check sum failure
Re-sold EEPROM or contact our Service.
The weighed articles are
overload .
Do not load the item exceeds the maximum
tolerance.
The calibration value
exceeds the full load
capacity while processing
single point Calibration.
Adjust the calibration value.
The accumulated number of
weighments, total count or
weight exceeds display
range.
No more accumulations.
(total count
window)
Under unit weight entering
or sampling mode, the
quantity of weighed articles
exceeds display range.
-----------------------------
(unit
weight window)
The calculate
unit weight
exceeds the display range
during sampling.
----------------------------
Low battery
Recharge the battery, the indicator can be
used while it is charging.

Questions and Answers: