Model 2500 Service Manual Calibration 2-7
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the
CAL pads which are located on the display board. Doing so will also reset the password to
the factory default (002500, front panel; KI002500, remote).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the
display board component layout drawing at the end of Section 6 for the location of the
CAL pads.
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be
accessible in the calibration menu. To view calibration dates and calibration count at any
time:
1. From normal display, press MENU, select CAL, and then press ENTER. The unit
will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2. Select VIEW-DATES and then press ENTER. The Model 2500 will display the
next and last calibration dates and the calibration count as in the following
example:
NEXT CAL: 04/15/2001
Last cal: 04/15/2000 Count: 0001

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals