4-2 Return to Section Topics 2600AS-900-01 Rev. B / September 2008
Section 4: Controlling Multiple Series 2600As (TSP-Link) Series 2600A System SourceMeter
®
Instruments User’s Manual
How do I set up the TSP-Link system?
Up to 32 TSP-Link enabled instruments (for example, Series 2600A System SourceMeter
TM
instruments) can be connected together to form a TSP-Link system.
Step 1: Connect the TSP-Link system
A connection example for an expanded system is shown in Figure 4-1. All the TSP-Link-enabled
instruments in the system are daisy-chained together using LAN crossover cables. In this
example, the unit connected to the PC is the Master and the other units are its Slaves. Note that
the Master does not have be assigned as Node 1, nor does it have to be at the end of the node
chain.
Figure 4-1
TSP-Link connections
Step 2: Assign node numbers
Each instrument in the TSP-Link system must have a unique node number assigned to it. The
node number for each instrument is stored in its nonvolatile memory and will not be lost when the
instrument is turned off.
Complete the following steps to assign a node number.
1. From the front panel, press the MENU > TSPLINK.
2. Select NODE and then use the navigation wheel to assign the node number.
Note: The node number can range from 1 - 64.
For remote programming, the tsplink.node attribute is used to set the node number for an
instrument:
tsplink.node = N
Where: N = 1 to 64. The node number of an instrument can be determined by reading the tsplink.node
attribute as follows:
nodenumber = tsplink.node
print(nodenumber)
The above print command will return the node number. For example, if the node number is 1, the
value
1.000000e00 will be returned.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL
B
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL
B
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL
B
!
Node 1 Node 2 Node 3 Node 32
Host
PC
LAN crossover cables
Type: Category 5e or higher.
Length: 3 meters maximum between nodes.
TSP-Link connectors
(2 per instrument)
NOTE The PC is not needed for
stand-alone systems.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL
B
!
LAN, RS-232
or GPIB

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611A and is the answer not in the manual?

Keithley 2611A Specifications

General IconGeneral
BrandKeithley
Model2611A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals