14 Quick Results Guide
Table 5
Auto ohms measurement procedure
Procedure Details
1. Select ohms measurement
function.
Press MEAS .
2. Select auto ohms measurement
method.
Press CONFIG > press MEAS > select
SOURCE > select AUTO.
3. Select measurement range. Use RANGE and keys to manually select
range. Or press AUTO RANGE (AUTO annunciator
on) to enable auto range.
4. Turn output on and take readings
from display.
Press ON/OFF OUTPUT key. OUTPUT indicator
on.
5. Turn output off when finished. Press ON/OFF OUTPUT key. OUTPUT indicator
off.
Table 6
Manual ohms measurement procedure
Procedure Details
1. Select ohms measurement
function.
Press MEAS .
2. Select manual ohms measurement
method.
Press CONFIG > press MEAS > select
SOURCE > select MANUAL.
3. Select source, set source level, and
set compliance limit.
See steps 1, 2, and 3 of Table 3.
4. Select measurement range (see
Notes 1 and 2).
Use RANGE and keys to manually select
range. Or press AUTO RANGE (AUTO annunciator
on) to enable auto range.
5. Turn output on and take readings
from display.
Press ON/OFF OUTPUT key. OUTPUT indicator
on.
6. Turn output off when finished. Press ON/OFF OUTPUT key. OUTPUT indicator
off.
Notes
1. If sourcing I, you will be selecting the V measure range. If sourcing V, you will be selecting the I measure
range.
2. The compliance range determines the maximum measurement range that can be selected.

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals