EasyManuals Logo
Home>Keithley>Measuring Instruments>6430

Keithley 6430 Quick Results Guide

Keithley 6430
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
26 Quick Results Guide
Logarithmic staircase sweep
This sweep is similar to the linear staircase sweep. The steps, however, are done on a
logarithmic scale. The symmetrical log points for the steps are determined by the specified
number of sweep points. Figure 11 shows a 5-point log sweep from 1 to 10V.
With the desired source (V or I) selected, use the following menu sequence to set the start
and stop levels, and the number of sweep points (2 to 2500):
Press CONFIG > press SWEEP > select TYPE > select LOG
After selecting LOG, you will be prompted to enter the START and STOP levels, and the
NO OF POINTS.
Delay
Delay
Delay
Delay
Start
Bias
Stop
(10)
Measure
#1
Measure
#2
Measure
#3
Measure
#5
X = Measurement Point
X
X
X
X
Delay
X
Measure
#4
Log Points = 5
1
10
5.6234
3.1623
1.7783
Volts
Log
Scale
Figure 11
Logarithmic staircase sweep (example)