EasyManuals Logo
Home>Keithley>Multimeter>DMM6500

Keithley DMM6500 User Manual

Keithley DMM6500
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
In this section:
Introduction .............................................................................. 6-1
Equipment required .................................................................. 6-1
Device connections .................................................................. 6-2
Measuring 4-wire resistance with offset compensation ............ 6-3
Introduction
This application example demonstrates how to use the DMM6500 to accurately measure resistance.
Typical resistance measurements made using the 2-wire method source current through the test
leads and the device under test (DUT). The voltage is measured and the resistance is calculated.
It is difficult to obtain accurate 2-wire resistance measurements when the DUT is lower than 100 Ω.
Typical lead resistances lie in the range of 1 mΩ to 10 mΩ. When the 2-wire method is applied to low-
resistance measurements, there is a small but significant voltage drop across the resistance of each
test lead. The voltage measured by the instrument is not the same as the voltage directly across the
DUT.
The 4-wire method is preferred for low-resistance measurements. With this configuration, the test
current is sourced through the DUT using one set of test leads, while a second set of SENSE leads
measures the voltage across the DUT. Connect the voltage-sensing leads as close to the DUT as
possible to avoid including the resistance of the test leads in the measurement.
Thermoelectric voltages (EMFs) can seriously affect low-resistance measurement accuracy. The
DMM6500 can apply the offset-compensated ohms method (OCOMP), which makes one normal
resistance measurement and one using the lowest current source setting to eliminate EMFs.
For this example, you use a 20 Ω resistor. Fixed measurement ranges are applied in order to optimize
scanning speed and OCOMP is applied to correct for any EMF effects.
For comprehensive information on 4-wire resistance measurements, thermoelectric EMFs, and offset
compensation methods, see the Low Level Measurements Handbook, available on tek.com/keithley
.
Equipment required
One DMM6500
One computer set up for communication with the instrument
Four insulated banana cables
One device to be tested (the application shown uses a 20 Ω resistor)
Section 6
Measuring 4-wire resistance with offset compensation

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
BrandKeithley
ModelDMM6500
CategoryMultimeter
LanguageEnglish

Related product manuals