Section
4: Making basic front-panel measurements Model 2460 Interactive SourceMeter® Instrument
User's Manual
4-2 2460-900-01 Rev. A / September 2014
Equipment required for this application
Equipment required for this application:
Model 2460 Interactive SourceMeter
®
instrument
Two insulated banana cables; you can use the set that is provided with the Model 2460, the
Keithley Instruments Model 8608 High-Performance Clip Lead Set
One 10 k resistor to test
Device connections
Turn the power to the instrument off before attaching connections to the Model 2460.
Connect the Model 2460 to the resistor in a 2-wire (local sense) configuration. In this configuration,
the device is connected between the FORCE HI and FORCE LO terminals.
The physical connections to the front panel are shown in the following figure.
Figure 17: Model 2460 2-wire front-panel connections
Make front-panel measurements
For this application, you will:
Select the source and measure functions
Select the source range
Set the source value
Set the source limit
Select the measurement range
Turn on the source output
Observe the readings on the display
Turn off the source output

Table of Contents

Other manuals for Keithley SourceMeter 2460

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2460 and is the answer not in the manual?

Keithley SourceMeter 2460 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals