18 FKB-BA-e-2010
Navigation key Press button, the first selectable adjustment weight will be
displayed.
Use the navigation keys  to select the desired adjustment weight, see chap. 1
„Adjustment points“ or „Recommended adjustment weight“
Prepare the required adjustment weight.
Acknowledge selection by button. < Zero >, < Pt ld > followed by the
weight value of the adjustment weight to be placed will be displayed.
Place the adjustment weight and confirm with button,
< wait > followed by < success > will be displayed.
After successful adjustment the balance automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing pan) the display will
show the error message < wrong >. Switch off balance and repeat the
adjustment process.
7.9.2 External adjustment with user-defined adjustment weight < caleud >
Observe stable environmental conditions. A warm up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing pan.
To invoke the setup menu press the TARE and ON/OFF button at the same time
and keep them pressed until the first menu item < Cal> will be displayed.
Navigation key Press button, < CalExt > will be displayed.
Use the navigation keys to select < caleud >.
Confirm with button. The numeric input window for the weight value of the
adjustment weight appears.
Enter weight value and confirm using the button, numeric entry see chap. 3.2.2.
< Zero >, < Pt ld > followed by the weight value of the adjustment weight to be
placed will be displayed.
Place the adjustment weight and confirm with button,
< wait > followed by < success > will be displayed.
After successful adjustment the balance automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing pan) the display will show
the error message < wrong >. Switch off balance and repeat the adjustment process.

Table of Contents

Other manuals for KERN FKB 8K0.1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN FKB 8K0.1 and is the answer not in the manual?

KERN FKB 8K0.1 Specifications

General IconGeneral
BrandKERN
ModelFKB 8K0.1
CategoryScales
LanguageEnglish

Related product manuals