24 FKB-BA-e-2010
The entered tare weight remains invalid until a new tare weight is input. To
delete press the TARE key or confirm the menu setting < clear > using
the TARE button.
10.1.2.2 Enter the known tare weight numerically < Ptare manuel >
Invoke menu setting < Ptare > and confirm by button.
Use the navigation keys to select the setting < manuel > and confirm with
button.
Enter the known tare weight, numeric input see chap. 3.2.2
The input weight is automatically saved as tare weight, the indicator (NET) and the
tare weight with minus sign will appear.
Place the filled weighing container.
Wait until the stability display appears ( ).
Read net weight.
The entered tare weight remains invalid until a new tare weight is input. To
delete enter
the zero value or confirm the menu setting <clear> using the
TARE button.
10.1.2.3 Data-Hold function < hold >
Invoke menu setting < hold >
Place goods to be weighed on balance.
Confirm with button.
The first stable weighing value is kept for 15 s, symbolised by [HOLD] in the upper
edge of the display.