35
Model CM-3600A
Illumination/
observation
system
Reflectance; di:8°, de:8° (diffused illumination, 8-degree viewing), equipped
with simultaneous measurement of SCI (specular component included) / SCE
(specular component excluded) Conforms to CIE No.15, lSO 7724/1, ASTM
E 1164, DIN 5033 Teil 7 and JIS Z 8722 condition c standard. Transmittance
: di:0°, de:0° (diffused illumination, 0-degree viewing) Conforms to CIE No.15,
ASTM E 1164, DIN 5033 Teil 7 and JIS Z 8722 condition g standard.
Light-receiving
element
Silicon photodiode array (dual 40 elements)
Spectral separation
device
Diffraction grating
Wavelength range 360 to 740 nm
Wavelength pitch 10 nm
HaIf bandwidth Approx.10 nm
Reflectance range 0 to 200%; resolution: 0.01%
Sphere size ø152 mm
Light source 4 pulsed xenon lamps
Minimum
interval between
measurements
Normal SCI/ SCE measurement: 4 sec.
Transmittance measurement: 3 sec.
UV-cut/ UV-adjusted measurement: 5 sec.
Measurement/
illumination area
(Selectable)
LAV : ø25.4 mm/ ø30 mm
MAV : ø8 mm/ ø11 mm
SAV : ø4 mm/ ø7 mm
Repeatability
When white calibration plate is measured 30 times at 10-sec.
intervals after white calibration has been performed;
Spectral reflectance: Standard deviation within 0.1%
Colorimetric values: Standard deviation within E
*
ab
0.02
Inter instrument
agreement
Mean E
*
ab
0.15 (LAV/SCI) Average for 12 BCRA Series II color
tiles compared to values measured with master body.
UV adjustment Instantaneous numerical adjustment
UV cut filter 400 nm cutoff and 420 nm cutoff
Transmittance
chamber
Width: 133 mm; depth: approx. 50 mm; measurement dia.: approx. 17 mm
Transmittance Specimen Holder (Optional accessory): Sample
holder for both plate-shaped and liquid samples (removable)
Interface USB 1.1
Power AC100 to 240 V 50/60 Hz (Using included AC adapter)
Operating
temperature/
humidity range (*1)
13 to 33°C, relative humidity 80% or less (at 33°C) with no condensation
Storage
temperature/
humidity range
0 to 40°C, relative humidity 80% or less (at 35°C) with no condensation
Size (WxHxD) 244 x 205 x 378 mm
Weight 11.5 kg
The specifications given here are subject to change without prior notice.
Specications

Table of Contents

Questions and Answers:

Konica Minolta CM-3600A Specifications

General IconGeneral
ModelCM-3600A
Light SourcePulsed Xenon lamp
Spectral Range360 nm to 740 nm
Wavelength Pitch10 nm
InterfaceUSB, RS-232C
TypeSpectrophotometer
Measurement Geometryd/8°
Integrating Sphere Diameter150 mm
Color SpacesHunter Lab, Munsell
Observer2°/10°
IlluminantD65, A, C
PowerAC 100-240V, 50/60 Hz
Measurement AreaMAV: Ø8 mm / SAV: Ø3 mm
RepeatabilityΔE*ab 0.03 (MAV, white tile)
Colorimetric DataCIE XYZ, CIE L*a*b*, CIE L*C*h

Summary

Safety Precautions

Using the CM-3600A

Measurement Procedure

Flow of Preparation and Measurement

Step-by-step guide for preparing and performing measurements with the CM-3600A.

Connecting a Personal Computer

Connecting the AC Adapter

Turning Power ON and OFF

Attaching a Target Mask

Notes on Use of Target Mask

Important notes regarding the handling and maintenance of target masks.

Attaching the Zero Calibration Box

Notes on Use of Zero Calibration Box

Guidelines for proper use and care of the zero calibration box.

Attaching the White Calibration Plate

Attaching the Transmittance Zero Calibration Plate (Optional Accessory)

Performing 100% calibration

Setting up instrument for 100% calibration to air

Procedure for setting up the instrument for 100% air calibration.

Setting up instrument for 100% calibration to water

Procedure for setting up the instrument for 100% water calibration.

Setting a Specimen

Reflectance measurement

Transmittance measurements

Cleaning the CM-3600A and Accessories

Error Message

TROUBLESHOOTING GUIDE

Illumination and Measurement Areas

Overview of Simultaneous Measurement of SCI/SCE

Fluorescent Measurement

WHEN FLUORESCENT CALIBRATION IS PERFORMED

Four calibration methods for accurate fluorescent reflectance measurement.

CALCULATING FLUORESCENT REFLECTANCE

How fluorescence is calculated using measured reflectances.

WHEN FLUORESCENT CALIBRATION IS NOT PERFORMED

Measuring fluorescent reflectance without specific calibration.

UV CUT LIGHT SOURCES

Details on available UV cut light sources for fluorescent measurements.