5.4 Environment Conditions
Use Conditions
Ambient Temperature 10°C - 40°C
Relative Humidity 30% - 75 %
Atmospheric Pressure 795 hPa - 1060 hPa
Storage and Transport Conditions
Ambient Temperature -20°C - 50°C
Relative Humidity 20% - 90 %
Atmospheric Pressure 795 hPa - 1060 hPa
5.5 Electrical Specications (Eleganza 4)
Input Voltage
Version 1
Version 2
230 V AC, 50/60 Hz
120 V AC, 50/60 Hz
Maximum Power Input 370 VAC
Ingress Protection (EN 60529) IP X4
Protection Class Class I
Electrical Motor Duty Cycle 2 minutes ON /18 minutes OFF
Accumulator Pb AKU 2 x 12 V / 1,2 Ah / Fuse 15 A
Fuse
Version 1
Version 2
2 x T1.6A L 250V for 230 V
2 x T3.15A L 250V for 120 V
NOTE Upon request, LINET ® can deliver hospital beds with electrical specications that comply with regional standards (custom
voltage, different mains plugs).
5.6 Electromagnetic Compatibility
Bed is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical system
for magnetic resonance imaging, where the intensity of EM disturbances is high.
Bed has dened no essential performance.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead to
improper operation. If such use is needed, this device and the other equipment should be under surveillance
to verify proper operation.
List of used cables:
► Mains cable, maximum length 6 m
► Attendant Control Panel, maximum length 3m
► Handset, maximum length 3m
WARNING!
Use of the accessories, converters and other cables, than specied and provided by manufacturer of this bed
could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this bed and lead
to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should
not be used closer than 30 cm (12 inches) from any part of this bed Eleganza 4, including cables specied by
manufacturer. Otherwise this could lead to deterioration of functionality of this bed.
WARNING!
Do not overload the bed (SWL), respect the duty cycle (INT.) and consider chapter 19 Maintenance in order to
maintain the basic safety with regard to electromagnetic disturbances for the expected service life.