13.5 PAUSE
During PAUSE Mode Bed Exit Monitoring is temporarily interrupted and Alarms are not activated.
When Outer Zone monitoring is activated, PAUSE period is terminated
when patient returns to the bed within 15 minutes of the PAUSE period.
To PAUSE Bed Exit Monitoring:
► Press button or button .
Icon 10 appears on the display with 15 minute countdown timer.
After PAUSE period elapsed and patient is in ordered position Bed Exit Monitoring is reactivated.
To extend the PAUSE period:
► Press button again to extend the countdown to 15 minutes period again.
To terminate the PAUSE period:
► Press button .
13.6 Deactivation of Bed Exit Monitoring
To deactivate Bed Exit Monitoring:
► Press button .
Icon appears on the display.
13.4 BED EXIT ALARM
Alarm is triggered when patient has left selected monitored zone or
PAUSE period elapsed and patient is not in ordered position.
To stop Alarm:
► Press button
.
Bed Exit Monitoring is deactivated and icon appears
on the display.
The audible alarm is stopped.
To pause Alarm:
► Press button
or button .
„Pause 15 min.“ appears on the display and 15 minute countdown timer runs.
The audible alarm is paused.
13.4.1 Alarm Volume
Maximum Alarm Volume Level is set by default.
It is possible to set Alarm Volume before and during triggered alarm.
To lower Alarm Volume Level:
► Press button
.
Icon with lower Alarm Volume Level appears on the display. Volume is lowered.
To return to Maximum Alarm Volume Level:
► Press button after Minimum Alarm Volume Level has been reached.
Icon with the 3 levels appears on the display.
1. Minimum Volume
2. Moderate Volume
3. Maximum Volume
321
Fig. Volume Icon
Fig. Visual signalisation of Bed Exit Alarm (iBoard Basic)
Fig. Bed Exit OFF (iBoard
Basic)
Fig. PAUSE Countdown
(with remaining minutes)