8.5 Before Use
Prepare the bed for use as follows:
Connect the bed to the mains.
Charge the accumulator.
Raise and tilt the mattress support platform to the highest position.
Lower and tilt the mattress support platform to the lowest position.
Check that the castors as well as main brake work correctly.
Check that the bed extension works correctly.
Check that it is possible to remove the head and foot boards.
Check all of the functions on the control elements.
Check that the siderails function properly.
Dispose of all packaging (see Disposal).
8.6 Transport
For a safe transport, observe the following:
Ensure that no cables are run over when moving a bed.
Ensure that the mains cable is attached with a hook (at the head end of the bed).
Ensure that the castors are unlocked before moving the bed during the loading/unloading process (see Castor Control).
Adjust bed height to at least 20 cm below maximum height.
Push bed by handles on head or foot end.
Move the bed exclusively on suitable oor surfaces.
For longer distances, ensure that the castor steering function is activated.
Ensure that the brakes are released while moving the bed.
Suitable surfaces:
Tile
Hard linoleum
Poured ooring
Unsuitable surfaces:
Too soft, unsealed or defective ooring
Soft wooden ooring
Soft and porous stone oors
Carpeted oors with underlay
Soft linoleum

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET WS17 and is the answer not in the manual?

LINET WS17 Specifications

General IconGeneral
BrandLINET
ModelWS17
CategoryMedical Equipment
LanguageEnglish

Related product manuals