MAXIMUM CAPACITY
: maximum displayable weight (from 0 to max full scale; default: 0). When the weight exceeds
this value by 9 divisions, the display shows . To disable this function, set 0.
TARE WEIGHT ZERO SETTING
This menu may also be accessed directly from the weight display, holding down the key
for 3 seconds.
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
- Confirm the message by pressing .
- The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing.
- Confirming once again, the weight is set to zero (the value is stored to the permanent memory).
- Press to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
DIAGNOSTICS ON ZERO: if diagnostics on zero has been enabled, the instrument can display and
store the load distributions on zero and the mV on active channels (see section TEST). If the stored
weight is zero and all channels are , it means that no zero-setting has yet been performed.
WARNING: diagnostics on zero is performed only if the load distribution has been stored
at least once.