- 36 -
ALARMS
: no load cell detected, check the connections.
: the load cell signal exceeds 39 mV; the conversion electronics (AD converter) is
malfunctioning.
: the load cell excitation is not connected or is incorrectly connected; the references are
not connected or are incorrectly connected; the load cell is a 4-wire and there are no
jumpers between EX- and REF- and between EX+ and REF+.
: the load cell is not connected or is incorrectly connected (the number indicates the
channel on which the error is detected).
: the weight display exceeds 110% of the full scale.
: internal instrument converter failure; check load cell connections, if necessary contact
technical assistance.
: the weight exceeds the maximum capacity by 9 divisions.
: maximum displayable value exceeded (value higher than 999999 or lower than -999999).
: weight too high: zero setting not possible.
: this message appears in the sample weight setting, in real calibration, after the fifth
sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press to quit the setting
mode leaving the previous value unchanged. Examples: a number of decimals is selected
for full scale which exceeds the instrument's display potential; value above the maximum
setting value; the weight value set in sample weight verification does not match the
detected mV increase; the analog output correction goes beyond the permitted limits.
: lock active on menu item, keypad or display.
: It’s not possible to display properly the number because is greater than 999999 or less
than -999999.
: the current load cell has already been equalized; press to go back to the previous
step and move the sample weight on the next load cell.
: the sample weight was not loaded or is too light.
: the load percentage error is higher than the value set in parameters  or ;
press to cancel the alarm.
: gross weight equal to zero: the semi-automatic tare operation cannot be performed.
: issues with the fieldbus device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Load Cell Systems TLB4 and is the answer not in the manual?

Load Cell Systems TLB4 Specifications

General IconGeneral
BrandLoad Cell Systems
ModelTLB4
CategoryTransmitter
LanguageEnglish