- 27 -
SERIAL COMMUNICATION SETTING
- : it disables any type of communication (default).
- : MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
- : ASCII bidirectional protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
- 
- 
- : continuous weight transmission protocol (see Communication protocols manual), at the
frequency set in  item (from 10 to 300).
-  (set:  = ,  = ).
-  (set:  = ,  = ).
- : continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series remote
displays; the remote display shows the net weight or gross weight according to its settings (set:
 = ,  = ,  = ).
- : continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series remote
displays; the remote display shows the net weight or gross weight according to its settings (set:
 = ,  = ,  = ).
- : continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series remote
displays (set:  = ,  = ,  = ).
When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross weight.
- if the instrument shows the net weight, the remote display shows the net weight alternated
with the message .
- : printer.
- : communication protocol with inclinometer (see Inclinometer manual).
- : transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
- : instrument address (from 1 to 99; default: 1).
- : maximum transmission frequency (10 20 30 40 50 60 70 80 100
200 300; default: 10); to be set when the  transmission protocol is selected.
Maximum setting frequency ():
- 20 Hz with minimum baud rate 2400 baud.
- 40 Hz with minimum baud rate 4800 baud.
- 80 Hz with minimum baud rate 9600 baud.
- 100 Hz with minimum baud rate 19200 baud.
- 200 Hz with minimum baud rate 38400 baud.
- 300 Hz with minimum baud rate 38400 baud.
- : delay in milliseconds which elapses before the instrument replies (from 0 to
200 ms; default: 0).


+

