- 32 -
SETPOINT PROGRAMMING
From the weight display, press to access the setpoint setting.
: to enter a menu/confirm the data entry.
: to modify the displayed figure or menu item.
: to select a new figure or modify the displayed menu item.
: to cancel and return to the previous menu.
- (from 0 to max full scale; default: 0): Setpoint; relay switching occurs when the weight
exceed the value set in this parameter. The type of switching is settable (see section OUTPUTS
AND INPUTS CONFIGURATION).
- (from 0 to max full scale; default: 0): Hysteresis, value to be subtracted from the setpoint
to obtain contact switching for decreasing weight. For example with a setpoint at 100 and
hysteresis at 10, the switching occurs at 90 for decreasing weight.
These values are set to zero if the calibration is changed significantly (see sections
THEORETICAL CALIBRATION and REAL CALIBRATION (WITH SAMPLE WEIGHTS)).