EasyManuals Logo
Home>Mindray>Measuring Instruments>BA-88A

Mindray BA-88A Operation Manual

Mindray BA-88A
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
6-4
This calibration method adopts five parameters:
0
R , a , b , c and d .
This calibration method requires at least five calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding
R
is equal to
0
R .
Parabola
Calibration formula: cbCaCR ++=
2
This calibration method adopts three parameters:
a , b and c .
This calibration method requires at least three calibrators. The calibration
parameters can be calculated through the method of polynomial least squares.
Spline
Calibration formula:
32
0
)()()(
iiiiiii
CCcCCbCCaRR +++=
This calibration method requires 2 to 6 calibrators. The number of calibrators is set
to be n, so the calibration method has 4(n-1) parameters in total:
i
R
0
,
i
a ,
i
b and
i
c .
6.3 Calculating Concentration
6.3.1 Calculating Concentration of Linearly Calibrated
Sample/control
One-point linear calibration
a
R
C
=
Where,
a - calibration parameter
Two-point linear calibration
a
bR
C
=
Where,
a , b - calibration parameters

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
BrandMindray
ModelBA-88A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals