NOTE
If either of the temperature sensor or analogue board is damaged, both
of the two should be replaced by one assembly at the same time. And
the replacement should be followed with temperature paremter
configuration.
During configuration, modifying parameters other than temperature
parameters should not be allowed. Otherwise, the test result might be
affected.
3.2.5 Base Plate Assembly
3.2.5.1 Introduction
The base plate assembly supports the system, providing fixing and support for
photometer assembly, hardware board, interface and other components.
3.2.5.2 Components and Structure
The basic structure of the base plate assembly is shown as figure 3-12.
Figure 3-12 Base plate assembly structure
3.2.5.3 Assembly and Disassembly of the Base Plate Assembly
The base plate assembly is a strutting piece of sheet metal, providing fixing and
support for photometer assembly, hardware board, interface and other components.
Pay attention to the following while assembling and disassembling it.
1 When assembling and disassembling the photometer assembly, do not drag the
cable with excessive force, especially the cable of the semiconductor