9-4 Structure and Assembly/Disassembly
9.3.1 Display (Monitor) Assembly
The disassembly tool: cross-headed screwdriver (M3, M4)
1. Unscrew two M4X12 screws, and remove the display cable cover;
2. Unscrew three M4 X 12 screws. Remove the power supply cable of the display and unplug the
signal cable.
3. Unscrew six M4X12 screws, and remove the display assembly upwards.
Operate the following procedures on the display assembly:
9.3.1.1 Monitors rear cover/adjustable keyboard/front cover assembly
1. Unscrew four M4X12 screws, and then remove the rear cover of the display.
NOTE:
The metal part is inserted in the hole of the plastic part. The damage of the display’s
rear cover may occur if removed by force.