12-10 Probes and Biopsy
Observe the graph here carefully to immerse the transducer. Only soak parts
of the transducer below the strain relief.
Repeated disinfection will eventually damage the probe, please check the
probe performance periodically.
Compatible Disinfectants
Please see Mindray Transducer Disinfectant Recommendation for the disinfectant use in
details.
12.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the transducer is in
good condition. After disinfecting the transducer, confirm that the transducer is in good
condition and store it in a suitable place so that the next examination can be conducted
smoothly.
1. To prevent the transducer from being damaged, do not store it in locations where it may
be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport the transducer under following ambient conditions:
Ambient temperature: -20°C ~ 55°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
3. Store and transport the transducer D6-2EA under following ambient conditions:
Ambient temperature: -10°C ~ 50°C
Relative humidity: 20% to 95% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
4. When the transducer is sent to MINDRAY Customer Service Department or sales
representative for repair, be sure to disinfect it and keep it in the carrying case to prevent
infection.
5. Disinfect the carrying case as necessary.