12 | ni.com | NI PXI-4130 Calibration Procedure
The setup in Figure 3 measures the 200 μA, 2 mA, 20 mA, and 200 mA current ranges of
channel 1. The setup in Figure 4 measures the 1 A current range of channel 0 and the 2 A current
range of channel 1.
Figure 3. Current Programming Accuracy Verification Setup for the 200 μA, 2 mA, 20 mA,
and 200 mA Current Ranges of Channel 1
Figure 4. Current Programming Accuracy Verification Setup for the 1 A Current Range of
Channel 0 and the 2 A Range of Channel 1
Table 7 lists the test points you must request from the NI 4130 in DC Voltage mode and measure
with the DMM to complete verification. For each iteration, set the current limit of the NI 4130
to the value in the Output column of Table 7.
To ensure the system has had adequate time to settle, wait one second after requesting a new
current limit before taking a measurement with the DMM.
When applicable, configure the Fluke 5700/5720A/5730A calibrator to the Resistance value for
the corresponding channel and range in Table 4. Enable external sense (4-wire mode) on the
NI 4130 Fluke 5700A/5720A/5730A
DMM
Voltage
Mode
+
GRD GND
HI
LO
SENSEOUTPUT
AUX
1
HI
LO
Auxiliary Power
Supply
NI APS-4100
+
AUX IN
NI 4130
+
LO Sense
HI Sense
HI
LO
DMM
Current
Mode
Auxiliary Power
Supply
+
AUX IN +
AUX IN
NI APS-4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXI-4130 and is the answer not in the manual?

National Instruments PXI-4130 Specifications

General IconGeneral
BrandNational Instruments
ModelPXI-4130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals