Page #7 background imageLoading...
Page #7 background image
- 7 -
Initial setting of the head type
Set the head type when the power is first turned ON after the Amplifier Unit is purchased.
JSetting procedure
When the power is first turned ON, the following is displayed.
Note: The head type does not need to be set from the second time that the power is turned ON.
Changing the head type
This feature is used to display the type of head to connect the
Amplifier Unit to.
Switch the SET/RUN switch to SET and change the head type by
following the procedure below.
JChange procedure
Select whether to turn output ON when the distance between workpiece and the Sensor Head is shorter than the threshold
value (near distance) or when it is longer than the threshold value (far distance).
The setting method varies depending on the Amplifier Unit model.
6. Basic Settings
1. Setting the Head Type
2. Setting the Operating Mode
Selection Details
NON (near ON)
(default)
Output turns ON when the distance between the workpiece and the Sensor Head is shorter than the threshold value.
FON (far ON) Output turns ON when the distance between the workpiece and the Sensor Head is longer than the threshold value.
Model Method
Twin output model The setting method is set by the "operating mode" in the SET mode.
Refer to "Function Transition Charts" in "5. Function Overview."
Analog output
model
Select this method by the operating mode switch.
The sub-display stops flashing and lights.
After one second, measurement starts.
Sub-display:Main display:
Select the model (30, 80, 100) by the UP or DOWN key,
and then press the MODE key to apply the setting.
(In this example, E4C-DS30(L) is selected.)
RU
N
SET
Select [8-SS] by the MODE key.
This com
p
letes settin
g
, and the dis
p
la
y
returns to re
g
ular measurement.
Switch the SET/RUN switch to RUN.
RUNSET
Select the model (30, 80, 100) by the UP or DOWN key,
and then press the MODE key to apply the setting.
(In this example, E4C-DS30(L) is selected.)
Sub-display:Main display:
FN
1636705-7A_e4c-uda_E.fm 7 ページ 2005年11月15日 火曜日 午後4時58分