22 . Lambda 365 Users Guide
General Operating Conditions
The Lambda 365 has been designed and tested in accordance with PerkinElmer specifications
and in accordance with the safety requirements of the International Electrotechnical
Commission (IEC). The Lambda 365 conforms to IEC61010-1 (Safety Requirements for
electrical equipment for measurement, control and laboratory use) as it applies to IEC Class 1
(earthed) appliances and therefore meets the requirements of EC directive 2006/95/EC.
Only use the Lambda 365 indoors and under the following conditions:
Temperature
l5
o
C to 35
o
C
R
elative Humidity Less than 80% relative non-condensing
Indoor only Altitude up to 2000 m
If possible, avoid any adjustment, maintenance and repair of the opened, operating
instrument. If any adjustment, maintenance and repair of the opened instrument is
necessary, this must be done by a skilled person who is aware of the hazard involved.
Whenever it is likely that the Lambda 365 is unsafe, make it inoperative. The Lambda 365
may be unsafe if it:
• shows visible damage
• fails to perform the intended measurement
• has been subjected to prolonged storage in unfavourable conditions
• has been subjected to severe transport stresses.
WARNING
If the equipment is used in a manner not specified herein, the
protection provided by the equipment may be impaired.
Si l'équipement est utilisé d'une manière non spécifiée par ceci, la
protection fournie par l'équipement risque d’être altérée.